Model
|
IRW-1000
|
IRW-2000
|
IRW-2000B
|
Operational Environment
|
Temperature within Operational Environment
|
+18°C~+30°C
|
+18°C~+30°C
|
+18°C~+30°C
|
Relative Humidity within Operational Environment
|
30°C~75%
|
30°C~75%
|
30°C~75%
|
Velocity of Air Flow within Operational Environment
|
<
0.3m/s
|
<
0.3m/s
|
<
0.3m/s
|
Power Requirement
|
≤
1000VA
|
≤
1000VA
|
≤
1000VA
|
Temperature Control Mode
|
Auto
|
Auto
|
Auto
|
Temperature Adjustment Range
|
25°C~37°C
|
25°C~37°C
|
25°C~37°C
|
Temperature Limit Override Mode
|
N/A
|
38°C
|
38°C
|
Temperature Sensor Accuracy
|
≤
0.3°C
|
≤
0.3°C
|
≤
0.3°C
|
Uniformity of Mattress Temperature
|
≤
2.0°C
|
≤
2.0°C
|
≤
2.0°C
|
Time to Warm Up
|
≤
45 min
|
≤
45 min
|
≤
45 min
|
Heat Output Indicate
|
0~100%
|
0~100%
|
0~100%
|
Transverse Rotation Range of Radiant Unit
|
0~±90° step-less
|
0~±90° step-less
|
0~±90° step-less
|
Mattress Tray Size
|
68cmx54cm
|
68cmx54cm
|
68cmx54cm
|
Mattress Tilt Adjustment Range
|
0~±10° step-less
|
0~±10° step-less
|
0~±10° step-less
|
Maximum Load of Mattress Tray
|
10kg
|
10kg
|
10kg
|
Maximum Load of IV Stand
|
2kg
|
2kg
|
2kg
|
Maximum Load of Shelf
|
2kg
|
2kg
|
2kg
|
X-ray Cassette
|
Yes
|
Yes
|
Yes
|
RS-232 Connector
|
N/A
|
Yes
|
Yes
|
Mattress of Radiant Unit
|
81cm
|
81cm
|
81cm
|
Heater Infrared Wavelength
|
1~3μm
|
1~3μm
|
1~3μm
|
APGAR Timer
|
N/A
|
Yes
|
Yes
|
Alarms
|
Power Failure Alarm
|
Yes
|
Yes
|
Yes
|
Sensor Failure Alarm
|
Yes
|
Yes
|
Yes
|
Temperature Deviation Alarm
|
±1.0°C
|
±1.0°C
|
±1.0°C
|
Temperature Exceeds Set Limit Alarm
|
≤
38.0°C
|
≤
38.0°C,
≤
39.0°C
|
≤
38.0°C,
≤
39.0°C
|
System Failure Alarm
|
N/A
|
Yes
|
Yes
|
Low-pressure Suction
|
N/A
|
N/A
|
Yes
|
Pressure Adjustment Range
|
N/A
|
N/A
|
0~22Kp (0~165mmHg)
|
Noise Level
|
N/A
|
N/A
|
≤55dB
|
White Phototherapy Light
|
N/A
|
N/A
|
1 set
|
7L Oxygen Cylinder (with Oxygen Manometer & Oxygen Flowmeter)
|
N/A
|
N/A
|
2 sets
|
Neonate Resuscitation Bag
|
N/A
|
N/A
|
1 set
|