Model
|
BI-3000
|
BI-1000
|
BI-800
|
Canopy Specifications
|
Mattress (canopy)
|
450mm
|
450mm
|
450mm
|
Access Port
|
6 (2 Iris ports)
|
6 (2 Iris ports)
|
4 (2 Iris ports)
|
Self-sealing Tube Ports
|
4
|
4
|
2
|
Front Access Panel
|
826x310mm
|
826x310mm
|
826x310mm
|
Mattress Tray Size
|
630x360mm
|
630x360mm
|
630x360mm
|
Technical Specifications
|
Temperature Mode (Servo Controlled)
|
air, infant
|
air, infant
|
air
|
Air Temperature Adjustment Range
|
20°C~37°C, 37~39°C
|
20°C~37°C, 37~39°C
|
20°C~37°C
|
Infant Temperature Adjustment Range
|
30°C~37°C, 37~38°C
|
30°C~37°C, 37~38°C
|
N/A
|
Infant Sensor Accuracy
|
≤
0.3°C
|
≤
0.3°C
|
N/A
|
Relative Humidity Display Range
|
5%~99%
|
N/A
|
N/A
|
Temperature Fluctuation
|
≤
0.5°C
|
≤
0.5°C
|
≤
0.5°C
|
Uniformity of Mattress Temperature
|
≤
0.8°C
|
≤
0.8°C
|
≤
0.8°C
|
Time to Warm Up
|
≤
30 min
|
≤
30 min
|
≤
30 min
|
Noise Level within Canopy
|
≤
55dB
|
≤
55dB
|
≤
55dB
|
Alarms
|
Power Failure Alarm
|
Yes
|
Yes
|
Yes
|
Fan Failure Alarm
|
Yes
|
Yes
|
Yes
|
Sensor Failure Alarm
|
Yes
|
Yes
|
Yes
|
Air Temperature Deviation Alarm
|
±3.0°C
|
±3.0°C
|
±3.0°C
|
Infant Temperature Deviation Alarm
|
±1.0°C
|
±1.0°C
|
N/A
|
Air Temperature Exceeds set Limit Alarm
|
≤
38.0°C;
≤
40.0°C
|
≤
38.0°C; 40.0°C
|
≤
38.0°C
|
Infant Temperature Exceeds set Limit alarm
|
≤
38.0°C;
≤
39.0°C
|
≤
38.0°C; 39.0°C
|
N/A
|
System Failure Alarm
|
Yes
|
Yes
|
N/A
|
Mute Alarm Function
|
Yes
|
Yes
|
N/A
|
Operational Environment
|
Temperature within Operational Environment
|
+20°C~+30°C
|
+20°C~+30°C
|
+20°C~+30°C
|
Relative Humidity within Operational Environment
|
5%~99%
|
5%~99%
|
5%~99%
|
Velocity of Air Flow within Operational Environment
|
<
0.3 m/s
|
<
0.3 m/s
|
<
0.3 m/s
|
Other Specifications
|
RS-232 connector
|
Yes
|
Yes
|
N/A
|
Tilt of Infant Bed (Adjustment Range)
|
±10° step-less
|
±10° step-less
|
N/A
|
Maximum Load of Mattress Tray
|
5kg
|
5kg
|
5kg
|
Maximum Load of IV Stand
|
2kg
|
2kg
|
2kg
|
Maximum Load of Cabinet
|
6kg
|
6kg
|
N/A
|